Showing all 5 results

ELITech

Microlab 300

In stock

SKU: PGA-MICROLAB300
Light Source

Quartz-iodine lamp 12V-20W.

Wavelength Range

Optical unit with 12 position filter wheel
340, 405, 505, 546, 578, 620 nm standard installed
6 positions for optional filters.

Photometric Range

0.1 to 2.3 Absorbance
Resolution 0.001 Abs.

Detector

Photo diode (320-1000 nm).

Blanking

Automatic zero setting.

Operator Interface

Membrane keyboard, for direct function and alpha-numeric entry
Optional external keyboard
High contrast graphical LCD display
Real time clock, 24 hours system.

Measurement Procedures

Kinetic measurement with linearity check
Kinetic, with linearity check and sample slope blank
Two point kinetic, with or without reagent blank
End point, with or without reagent blank
Bichromatic end point, with or without reagent blank
End point, with sample blank and with or without reagent blank.

Multiple Testing

Up to nine replicates
Means, SD and CV.

Measurement Time

Programmable, 2 to 998 seconds for kinetic and two point type of tests
For end point fixed at 2 seconds.

Delay Time

Programmable, 0 to 999 seconds.

Parameter Setting

Method name
Measurement mode
Wavelength 1 and 2
Aspiration volume
Measurement delay
Measurement time
Factor
Concentration standards
Reagent blank y/n
Sample blank y/n
Units for results
Levels for flagging
Curve fit y/n
Linearity check.

Calibration

Factor, one-point, two-point and multi-point
Automatic on 1 standard (linear mode)
Automatic on up to 10 standards (non-linear mode).

Quality Control

Two controls per test
QC survey of last 30 control measurements
Levey Jenning plot
High/low flags.

Flowcell

Metal, with quartz windows, measuring volume 30 µL.

Temperature Control

By means of Peltier elements
Fixed temperature at 37°C.

Aspirator System

Internal pump of bellows type, driven by stepper motor:
Back panel connection for waste
Aspiration volume programmable.

Printer

Internal matrix printer
Normal paper
External printer port available.

Signal Interface

Centronics type parallel port
RS 232 type serial port
PS 2 type port for external keyboard.

Standards and Regulations

CE -IVD
CB
UL.

Dimensions and Power Requirements

40 x 17 x 36,6 cm (W x H x D)
Weight: 8,5 kg.
100-240 VAC nominal, 50/60 Hz
Battery back-up to retain data

Selectra Pro M

In stock

SKU: PGA-SELECTRAPRO-M
Kecepatan

Sampai dengan 266 tes per jam
Sampai dengan 180 tes per jam pada mode MONO

Sumber Cahaya

Lampu Quartz-iodine 12V-20W

Fotometrik

Absorbansi – 0.1 hingga 3.0;
Resolusi 0,001 Abs.

Konsumsi Air

Maksimal ~500 ml per jam, dengan operasi terus menerus

Standard dan Regulasi

ce-ivD
cb

Dimensi

122 x 75 x 61 cm (P x T x L)

Berat

95kg

Antarmuka

Host: RS232 atau Etherhet (TCP/IP) melalui protokol LIS-2A.

Selectra Pro XL

In stock

SKU: PGA-SELECTRAPRO-XL
Throughput

Up to 450 photometric and electrolyte tests/hour.

Reagent System

64 refrigerated reagent positions for 10 mL, 25 mL and 50 mL reagent bottles
Cooled to 10°C +/- 4°C at normal laboratory conditions
All reagent positions can be assigned as R1, R2 and R3.

Sample System

80 barcode readable positions
All positions fit 13×75 or 16x 100mm primary and secondary tubes and pediatric cups
All positions can be used for calibrators, controls, normal, pediatric and STAT samples

Pipetting System

1000 µL reagent syringe:
R1 volume 110-400 µL
R2 volume 0-180 µL
R3 volume 0-180 µL
Programmable in 1 µL steps
100 µL sample syringe:
Sample volume 1-30 µL
Programmable in 0.1 µL steps
Pre programmed ratio selectable from ratios 1:2 up to 1:100 with 3 diluents
Pre-heated probe with level detection, collision protection and integrated mixer.

Cuvette Rotor

Cost effective, semi-disposable cuvette rotor with 48 cuvettes
Up to 10.000 tests per rotor
Typical reagent consumption 220 µL per test
Measurement temperature 37°C, controlled by Peltier elements.

Light Source

Quartz-iodine lamp 12V-20W

Wavelength Range

Optical unit with 8 position filter wheel
340, 405, 505, 546, 578, 620, 660, 700 nm standard wavelengths
Other wavelengths available on request.

Photometric Range

-0.1 to 3.0 Absorbance
Resolution 0.001 Abs.

Analytical Modes (Single, Dual and Triple reagent system)

Kinetic measurement with linearity check
Mono- and bichromatic end point measurement with or without bichromatic reagent blank and/or sample blank correction
Two point measurement; with or without slope blank
Graphic plot of all measuring points
Predilution and automatic reflex dilution as needed
Range of calibration models
Non-linear calibration curves
Prozone check for immunology tests
Cut-off declaration
Calculated tests.

Quality Control

Up to 15 controls can be defined, 3 per test
Westgard or Basic QC rules
Levey-Jennings plots
Quality control statistics.

Standards and Regulations

CE -IVD
CB
UL.

Dimensions

117 x 115 x 80 cm (W x H x D)
46.1 x 45.3 x 31.5 in(W x H x D)

Weight

190 kg (419 lbs).

Interface

Host-Query interface available
Host: RS 232 or Ethernet (TCP/IP) through LIS-2A protocol.

Environmental Conditions

Temperature: 15 – 32 °C (59 – 90 °F)
Humidity: 15 – 88% RH
Altitude: up to 2000 m
Plumbing: No dedicated water or drain required
Electrical:
Voltage: 100 – 240 Vac
Frequency: 50 – 60 Hz
Power (max): 380 VA .

Integrated PC

Touch screen navigation
Operating system: Windowsâ„¢ XP

Barcode Reader

Hand held CCD barcode reader used for reagent identification

Eco Kit

Enables separate collection of concentrated measurement mixture and dilute (wash) waste

Dry ISE Module

Patented Solid State Electrode Technology
Indirect measurement
Dilution 1:14
Measures Sodium, Potassium, Chloride and CO2.

Positive Sample Identification

Positive Sample Identification (PSID) reads all popular barcodes.

Printer

Printer supported by MS Windowsâ„¢.

ProActive Maintenance Kit

Complete parts kit for annual preventive maintenance.

Selectra Pro XS

In stock

SKU: PGA-SELECTRAPRO-XS
Throughput

Up to 75 tests per hour.

Reagent System

Inner rotor ring:
15 reagent positions for 10 ml, 25 ml and 50 ml reagent bottles
Optionally cooled to 10°C +/ 4°C at normal laboratory conditions.
Outer rotor ring:
25 positions for (STAT) samples, controls and calibrators
15 positions for reagent bottles
All positions fit 13×75 mm primary and secondary tubes and pediatric cups
All positions can be used for calibrators, controls, normal, pediatric and STAT samples.

Pipetting System

100 μl reagent syringe:
– R1 volume 110-400 μl
– R2 volume 0-180 μl
-R3 volume 0-180 μl
– Programmable in 1 μl steps
100 μl sample syringe:
– Sample volume 1-30 μl
– Programmable in 0.1 μl steps
– Programmable dilution ratios 1:5 up to 1:100 in one step increments with 3 possible diluents
Pre-heated probe with level detection, collision protection and integrated mixer.

Cuvette Rotor

Disposable cuvette rotor with 48 cuvettes
Minimum measurement volume 220 μl
Measurement temperature 37°C, controlled by Peltier elements.

Light Source

Quartz-iodine lamp 12V-20W.
Wavelength Range
Optical unit with 8 position filter wheel
Automatic wavelength selection
340, 405, 505, 546, 578, 620, 660 nm standard installed
One free position
Other wavelenghts available on request.

Photometric Range

-0.1 to 3.0 Absorbance
Resolution 0.001 Abs.

Analytical Modes (Single, Dual and Triple reagent system)

Kinetic measurement with linearity check
Mono and bichromatic end point measurement with or without bichromatic reagent blank and/or sample blank correction
Two point measurement; with or without slope blank
Graphic plot of all measuring points
Automatic rerun with sample reduction
Non-linear calibration curves
Prozone check for immunology tests
Cut-off declaration
Calculated tests.

Quality Control

Up to 15 different controls can be defined, 3 per test
Westgard rules
Levey-Jennings plots
Quality control statistics.

Standards and Regulations

CE -IVD
CB.

Dimensions

89 x 58 x 60 cm (W x H x D)
No peripherals

Weight

67 kg.

Interface

State of the art Host-Query interface available
Host: RS 232 or Ethernet (TCP/IP) through LIS-2A protocol.

Environmental Conditions

Temperature: 15 – 32 °C
Humidity: 15 – 85% RH
Altitude: up to 2,000 m
Plumbing: No dedicated water or drain required
Electrical:
– Voltage: 100 – 240 Vac
– Frequency: 50 – 60 Hz
– Power (max): 350 VA .

Barcode Reader

Hand held CCD bar code reader used for test requisition, reagent identification and automated programing of assays, controls, and calibrators.

Printer

Printer supported by MS Windowsâ„¢

ProActive Maintenance Kit

Complete parts kit for annual preventive maintenance.