Showing 1–12 of 13 results

Alat Kimia Klinik adalah instrumen yang digunakan dalam laboratorium medis untuk menganalisis sampel biologis seperti darah, urin, dan cairan tubuh lainnya.

Tujuan utama dari alat kimia klinik ini adalah membantu dalam diagnosis, pemantauan, dan penelitian berbagai kondisi kesehatan, termasuk penyakit metabolik, infeksi, serta gangguan organ tubuh.

Alat Kimia Klinik

Chempro 3.0

In stock

SKU: PGA-CHEMPRO3.0
Detection Method

Photometric detection

Detection time

<3mins

CV

<10%

Sample Type

Serum, whole blood, plasma

Sample volume

45uL

Power supply

Adapter input AC100~240V,50/60Hz, 30VA: Output 5V,2A

Built-in Lithium battery

DC 3.7 +- 0.3V

Measurement

157.5*145.7*64.6mm

Weight

622g

Data storage

3000 results

Display

4.3' touch screen

Storage environment

Temp 20C ~ 60C; Rh: 10% ~ 90%

Operation environment

Temp 15C ~ 35C; Rh:<85%

Chemray 120

In stock

SKU: PGA-CHEMRAY120
Throughput

Lebih dari 100 test / jam.

Konten Pengujian

Clinical Chemistry, Immunoturbidimetric.

Mode Analisa

End-point, Two-point, Kinetic.

Programming

Open system with user defined programs.

Otomatis

Akses Acak.

Posisi Sampel

9 posisi untuk sampel.

Ukuran Sampel

3 ~ 45 uL, 0,5 uL adjustable.

Posisi Reagent

26 Posisi untuk Reagent.

Ukuran Reagent

R1 : 180 ~ 450 uL, 1 uL adjustable.
R2 : 3 ~ 450 uL, 1 uL adjustable.

Posisi Reaksi

90 cuvettes, 10 cuvettes/strip.

Cuvette

Optical Length 5mm.

Ukuran Reaksi

180 ~ 500ul.

Temperatur Reaksi

37C, Fluctuation 0.3C.

Print Out

Built in thermal printer, external printer optional.

Chemray 240

In stock

SKU: PGA-CHEMRAY240
Throughput

Lebih dari 160 test / jam.

Konten Pengujian

Clinical Chemistry, Immunoturbidimetric.

Mode Analisa

End-point, Two-point, Kinetic.

Programming

Open system with user defined programs.

Otomatis

Akses Acak.

Posisi Sampel

40 posisi untuk sampel.

Ukuran Sampel

3 ~ 100 uL, 0,5 uL adjustable.

Posisi Reagent

40 Posisi untuk Reagent.
Refrigerated Temperatre : 16C

Ukuran Reagent

R1 : 180 ~ 450 uL, 1 uL adjustable.
R2 : 30 ~ 450 uL, 1 uL adjustable.

Posisi Reaksi

81 cuvettes, 9 cuvettes/strip.

Cuvette

8 Step Washer System.

Ukuran Reaksi

180 ~ 500ul.

Temperatur Reaksi

37C, Fluctuation 0.2C.

Glory 917 (TERBARU 2024)

In stock

SKU: PGA-GLORY917-2024
Machine type

Random access, open reagent & locked reagent (optional)

Test Speed

Constant 150 tests/hour (Mono reagent or double reagent)

Test Principle

Colorimetric method, turbidimetry

Test Method

1 point end, 2 point end, Fixed-Time, Kinetics

Calibration Type

Linear & nonlinear

Sample Tray

40 sample positions

Sample Tube

Micro cup & test tube & blood collection tube can be used. ( Φ 12 -13 )mm * ( 25 – 100 ) mm

Sample Volume

2-30μL, step by 0.1μL

Sample Probe

Liquid surface detection, timely tracking for liquid volume.
Collision avoidance

Sample Barcode

Sample barcode reader (optional)

Reagent Tray

80 reagent positions (40 for R1 & 40 for R2), Non-stop cooling system with peltier pad inside, 24 hours 2°C-14°C, open system, could accept any reagent.

Reagent Volume

20-300μl, step by 1μl

Reagent Needle

Liquid surface detection, timely tracking for liquid volume.
Collision avoidance.

Regent Barcode

Regent barcode reader (optional)

Reaction tray

44 reaction cuvettes, new amorphous optical plastics reaction cups, perfect transmittance.

Reaction Volume

150μl-330μl

Reaction Time

10 minutes

Temperature

Peltier pad incubation system, reaction temperature (37℃ ± 0.1℃)

Mixer

Paddle-type mixing, effective reducing of carry-over

Light Source

Halogen lamp

Wavelength

340nm; 405nm; 450nm; 505nm; 546nm; 578nm; 630nm; 700nm(4 more options)

Absorbance range

0-4.0 Abs

Resolution

0.0001 Abs

Operation System

Windows XP, Windows 7, Windows 8, Windows 10

Data Processing

Automatic calibration, test combination, effective time management of reagents, serum index, whole process detection, linear expansion of enzyme, blank deduction,dirty cuvette memory , cross infection prevention, patient information memory and association input, automatic report audit, data fuzzy query, report statistics and printing, reference range grading, alarm information classification, user operation privilege classification management.

Report Printing

6 formats optional, surpport custom pattern

Interface

TCP/IP network interface

Power Supply

AC 110/220V ± 10%, 50/60 Hz, 650W

Water Consumption

6L/hour(maximum)

Dimension

750mm * 560mm * 535mm(L*W*H)

Weight

52 kg

PUSHKANG MS100

In stock

SKU: PGA-PUSHKANGMS100
Sample type

Anticoagulant whole blood, plasma, serum, capillary blood

Test principle

Photoelectric colorimetry

Connection

On-line of LIS system

Sample Volume

140μL

Test time

≦15min

Code identification

Automatic identification of QR code for reagent disc

Reagent type

Lyophilized reagent

Light source

LED lamp

Printer

Built-in thermal printer and external printer (wireless printer driver)

Operating system

Android 5.1 (5-inch)

Memory capacity

1GB, with expandable 32GB TF card

Supply voltage

AC 100V-240V, 50-60HZ

Size

270mm*180mm*155mm

Weight

3 KG, with built-in adapter

Selectra Pro M

In stock

SKU: PGA-SELECTRAPRO-M
Kecepatan

Sampai dengan 266 tes per jam
Sampai dengan 180 tes per jam pada mode MONO

Sumber Cahaya

Lampu Quartz-iodine 12V-20W

Fotometrik

Absorbansi – 0.1 hingga 3.0;
Resolusi 0,001 Abs.

Konsumsi Air

Maksimal ~500 ml per jam, dengan operasi terus menerus

Standard dan Regulasi

ce-ivD
cb

Dimensi

122 x 75 x 61 cm (P x T x L)

Berat

95kg

Antarmuka

Host: RS232 atau Etherhet (TCP/IP) melalui protokol LIS-2A.

Selectra Pro XL

In stock

SKU: PGA-SELECTRAPRO-XL
Throughput

Up to 450 photometric and electrolyte tests/hour.

Reagent System

64 refrigerated reagent positions for 10 mL, 25 mL and 50 mL reagent bottles
Cooled to 10°C +/- 4°C at normal laboratory conditions
All reagent positions can be assigned as R1, R2 and R3.

Sample System

80 barcode readable positions
All positions fit 13×75 or 16x 100mm primary and secondary tubes and pediatric cups
All positions can be used for calibrators, controls, normal, pediatric and STAT samples

Pipetting System

1000 µL reagent syringe:
R1 volume 110-400 µL
R2 volume 0-180 µL
R3 volume 0-180 µL
Programmable in 1 µL steps
100 µL sample syringe:
Sample volume 1-30 µL
Programmable in 0.1 µL steps
Pre programmed ratio selectable from ratios 1:2 up to 1:100 with 3 diluents
Pre-heated probe with level detection, collision protection and integrated mixer.

Cuvette Rotor

Cost effective, semi-disposable cuvette rotor with 48 cuvettes
Up to 10.000 tests per rotor
Typical reagent consumption 220 µL per test
Measurement temperature 37°C, controlled by Peltier elements.

Light Source

Quartz-iodine lamp 12V-20W

Wavelength Range

Optical unit with 8 position filter wheel
340, 405, 505, 546, 578, 620, 660, 700 nm standard wavelengths
Other wavelengths available on request.

Photometric Range

-0.1 to 3.0 Absorbance
Resolution 0.001 Abs.

Analytical Modes (Single, Dual and Triple reagent system)

Kinetic measurement with linearity check
Mono- and bichromatic end point measurement with or without bichromatic reagent blank and/or sample blank correction
Two point measurement; with or without slope blank
Graphic plot of all measuring points
Predilution and automatic reflex dilution as needed
Range of calibration models
Non-linear calibration curves
Prozone check for immunology tests
Cut-off declaration
Calculated tests.

Quality Control

Up to 15 controls can be defined, 3 per test
Westgard or Basic QC rules
Levey-Jennings plots
Quality control statistics.

Standards and Regulations

CE -IVD
CB
UL.

Dimensions

117 x 115 x 80 cm (W x H x D)
46.1 x 45.3 x 31.5 in(W x H x D)

Weight

190 kg (419 lbs).

Interface

Host-Query interface available
Host: RS 232 or Ethernet (TCP/IP) through LIS-2A protocol.

Environmental Conditions

Temperature: 15 – 32 °C (59 – 90 °F)
Humidity: 15 – 88% RH
Altitude: up to 2000 m
Plumbing: No dedicated water or drain required
Electrical:
Voltage: 100 – 240 Vac
Frequency: 50 – 60 Hz
Power (max): 380 VA .

Integrated PC

Touch screen navigation
Operating system: Windows™ XP

Barcode Reader

Hand held CCD barcode reader used for reagent identification

Eco Kit

Enables separate collection of concentrated measurement mixture and dilute (wash) waste

Dry ISE Module

Patented Solid State Electrode Technology
Indirect measurement
Dilution 1:14
Measures Sodium, Potassium, Chloride and CO2.

Positive Sample Identification

Positive Sample Identification (PSID) reads all popular barcodes.

Printer

Printer supported by MS Windows™.

ProActive Maintenance Kit

Complete parts kit for annual preventive maintenance.

Selectra Pro XS

In stock

SKU: PGA-SELECTRAPRO-XS
Throughput

Up to 75 tests per hour.

Reagent System

Inner rotor ring:
15 reagent positions for 10 ml, 25 ml and 50 ml reagent bottles
Optionally cooled to 10°C +/ 4°C at normal laboratory conditions.
Outer rotor ring:
25 positions for (STAT) samples, controls and calibrators
15 positions for reagent bottles
All positions fit 13×75 mm primary and secondary tubes and pediatric cups
All positions can be used for calibrators, controls, normal, pediatric and STAT samples.

Pipetting System

100 μl reagent syringe:
– R1 volume 110-400 μl
– R2 volume 0-180 μl
-R3 volume 0-180 μl
– Programmable in 1 μl steps
100 μl sample syringe:
– Sample volume 1-30 μl
– Programmable in 0.1 μl steps
– Programmable dilution ratios 1:5 up to 1:100 in one step increments with 3 possible diluents
Pre-heated probe with level detection, collision protection and integrated mixer.

Cuvette Rotor

Disposable cuvette rotor with 48 cuvettes
Minimum measurement volume 220 μl
Measurement temperature 37°C, controlled by Peltier elements.

Light Source

Quartz-iodine lamp 12V-20W.
Wavelength Range
Optical unit with 8 position filter wheel
Automatic wavelength selection
340, 405, 505, 546, 578, 620, 660 nm standard installed
One free position
Other wavelenghts available on request.

Photometric Range

-0.1 to 3.0 Absorbance
Resolution 0.001 Abs.

Analytical Modes (Single, Dual and Triple reagent system)

Kinetic measurement with linearity check
Mono and bichromatic end point measurement with or without bichromatic reagent blank and/or sample blank correction
Two point measurement; with or without slope blank
Graphic plot of all measuring points
Automatic rerun with sample reduction
Non-linear calibration curves
Prozone check for immunology tests
Cut-off declaration
Calculated tests.

Quality Control

Up to 15 different controls can be defined, 3 per test
Westgard rules
Levey-Jennings plots
Quality control statistics.

Standards and Regulations

CE -IVD
CB.

Dimensions

89 x 58 x 60 cm (W x H x D)
No peripherals

Weight

67 kg.

Interface

State of the art Host-Query interface available
Host: RS 232 or Ethernet (TCP/IP) through LIS-2A protocol.

Environmental Conditions

Temperature: 15 – 32 °C
Humidity: 15 – 85% RH
Altitude: up to 2,000 m
Plumbing: No dedicated water or drain required
Electrical:
– Voltage: 100 – 240 Vac
– Frequency: 50 – 60 Hz
– Power (max): 350 VA .

Barcode Reader

Hand held CCD bar code reader used for test requisition, reagent identification and automated programing of assays, controls, and calibrators.

Printer

Printer supported by MS Windows™

ProActive Maintenance Kit

Complete parts kit for annual preventive maintenance.